Patient Range
|
Pediatric(10
kg)-Adult |
|
Control
Settings
|
|
|
|
|
|
Modes (Pressure
or Volume): |
A/CMV
SIMV
SPONT |
|
Volume Control
(Tidal volume): |
100 to 2,200
mL |
|
(Frequency): |
1 to 99b/min |
|
tl
(Inspiratory Time): |
0.1 to 3.0 sec |
|
PEEP/CPAP: |
0 to 30 cmH2O/0
to 30 mbar |
|
Psupport (Pressure
Support): |
0 to 60 cmH2O/0
to 60 mbar* |
|
(Flow): |
6 to 100 L/min |
|
I:E Ratio: |
1:99 to 3:1 |
|
Pressure Control
(Target Pressure): |
5 to 60 cmH2O/5
to 60 mbar |
|
P trig (Sensitivity): |
-9.9 to 0 cmH2O/-9.9
to 0 mbar |
|
Panel Lock: |
ON/OFF |
|
Manual Inflation: |
3 sec maximum
(single operator controlled breath) |
|
FIO2: |
.21 to 1.00 |
|
Humidifier (Optional):
|
19°C to
39°C |
|
|
|
|
Monitor Parameters
| |
|
Paw (Peak, Mean,
Base): |
VT (insp.
tidal volume) |
|
Proximial Airway
Temperature: |
(insp. minute volume) |
|
Internal Battery: |
total
(total frequency) |
|
|
|
|
Alarms
|
|
|
Paw (High Airway Pressure): |
4 to 99 cmH2O/4
to 99 mbar |
|
Paw (Low Airway Pressure): |
3 to 98 cmH2O/3
to 98 mbar |
|
High Baseline
Pressure: |
Automatically
set |
|
Low Baseline
Pressure: |
Automatically
set |
|
l
(High Insp. Minute Volume): |
1.3 to 50 L |
|
l
(Low Insp. Minute Volume)
Back-up Ventilation: |
.3 to 49.0 L |
|
Apnea (Paw/Apnea): |
30 sec |
|
Device Alert: |
malfunction
alert |
|
Power Switchover: |
AC to DC |
|
Humidifier: |
humidifier fail/high
temp |
|
Low Battery: |
approx. 2 hours
battery remains |
|
Silence/Reset: |
60 sec. |
|
*PEEP + Psuppport cannot exceed 60 cmH2O/mbar |
|
|
|
|
|
|
|
Pneumatics
|
|
Gas
Delivery System: |
internal
dual micro-piston |
|
Air/Oxygen
Mixer (Optional): |
35
to 90 psig oxygen |
|
|
|
|
Power
Requirements |
|
|
|
100-240
VAC, max. 2A
50/60/400 Hz
12-30 VDC, max. 12A |
|
|
|
|
Internal
Battery |
|
|
|
Sealed
lead acid, 12V 5AH
provides up to 10 hrs of
operation time |
|
|
|
|
Dimensions
|
|
|
Width: |
10.63
in (27 cm) |
|
Depth: |
7.87
in (20 cm) |
|
Height: |
10.24
in (26 cm) |
|
Weight: |
15lbs
(6.8kg) (w/o humidifier) |
|
|
17.2lbs
(7.8kg)(with humidifier) |
|
|
|
|
Miscellaneous |
|
|
|
|
|
Operating
Temperature: |
-18°C
to 50°C |
|
Operating
Altitude: |
0
to 15,000 ft (0 to 4,572m) |
|
Storage
Temperature: |
-40°C
to 65°C |
|
RS-232/RS-485: |
8
pin Semconn connector
19,200 Baud |
|
|
|
|
Ordering
Information |
|
|
|
|
|
HT50-H: |
Ventilator
with integrated humidifier |
|
HT50-H1: |
Ventilator
(without integrated humidifier) |
|
HT56231: |
Compact
stand with support arm |
|
HT460700: |
Air/Oxygen
Entrainment Mixer |
|
HT600044: |
Oxygen
Blending Bag Kit |
|
HT460085: |
Automotive
lighter 12V Power cord |
|
HTECH01: |
E-cylinder holder
for stand |
|
|
|