The new Total Synchrony SAVI¢â. Even the
smallest infants can now breathe easier.
Ventilation
Application |
|
Neonatal |
|
|
Pediatric
|
CONTROLS
|
|
|
Modes
of Ventilation |
|
OFF
|
|
|
CPAP |
|
|
VENT
(Pressure Assistance* IMV,CMV) |
|
|
*Available
with SAVI |
Flow |
|
0
to 32 LPM; Flush 40 LPM |
Oxygen
Percent |
|
21%
to 100% |
Inspiratory
Time |
|
0.10
to 2.90 sec. |
Expiratory
Time |
|
0.30
to 60.0 sec. |
Manual
Breath |
|
|
Inspiratory
Pressure |
|
5
to 70 cm H2O |
Expriatory
Pressure |
|
-2
to 20 cm H2O |
Test |
|
Checks
microprossor function, displays and alarm function |
|
|
Waveform
Control |
|
|
Allows
the user to vary the proximal airway pressure waveform |
Time
Preset |
|
While
in the CPAP mode, this feature allows the user to preview
and select the desired Inspiratory and Expiratory Times
before switching to the VENT mode. |
Safety
Pressure Relief Valve (Adjustable) |
|
15
to 85 cm H2O |
ALARM
CONTROLS |
|
|
Delay
Time |
|
3
to 60 sec. |
|
|
Sets
the time delay before the LO pressure alarm is audibly
and visually activated. |
Alarm
Mute |
|
30
± 5 sec. |
Alarm
Set Arrows |
|
High
Inspiratory Pressure Alarm Limit |
|
|
Low
Inspiratory Pressure Alarm Limit |
Select
(Reset) Button |
|
Allows
the user to select either the LO or HI pressure alarm
limit setting. This button is also used to cancel the
HI limit alarm when it is activated. |
Alarms |
|
High
(HI) Pressure Limit |
|
|
Peak
Overpressure |
|
|
Low
(LO) Pressure Limit |
|
|
Low
Airway Pressure |
|
|
Leaks |
|
|
Patient
Disconnect |
|
|
Fail-to-Cycle |
|
|
Apnea |
|
|
Prolonged
Inspiration |
|
|
Source
Gas Failure |
|
|
Power
Failure (Battery Backup) |
Displays |
|
Inspiratory
Time |
|
|
Expiratory
Time |
|
|
I:E
Ratio |
|
|
Rate |
|
|
Airway
Pressure (to 80 cm H2O) |
|
|
High
Pressure Alarm |
|
|
Low
Pressure Alarm |
|
|
Mean
Airway Pressure |
Indictators |
|
Inspiratory
Phase |
|
|
Inverse
I:E Ratio |
|
|
Mode
Selection |
|
|
Alarm |
|
|
High
and Low Pressure Alarm |
|
|
Limit
Set Points |
Pre-set |
|
Safety
Auto Lock-Out Circuit (Inspiratory Phase terminated
at 4 sec. Even if microprossor fails |
Pneumatic
Requirements |
|
Medical
Grade Air and Oxygen: 30 to 60 psi; differential within
20 psi. |
Electrical
Requirements |
|
117
VAC 50/60 Hz |
|
|
Overseas
voltage requirements available) |
|
|
Circuit
Breaker Protected |
|
|
D.C.
Power Requirements |
|
|
12
VDC 1.2 amps |
Physical
Specifications |
|
|
Dimensions
|
|
Height:
16.25 in (41.27 cm) |
|
|
Width:
12 in (30.48 cm) |
|
|
Depth:
9.5 in (24.13 cm) |
Weights |
|
Net:
27 lbs. (12.2 Kg) |
|
|
Shipping:
35 lbs. (15.9 Kg) |
Miscellaneous
Specifications |
|
3-Wire
SJT Power Cord |
|
|
Hospital-Grade
Plug, current leakage less than 20 microamps when tested
per UL 544 |
|
|
Auxiliary
Flowmeter Port |
|
|
SAVI
Interface Cable |
|
|
Specifications
subject to change without notice |
|
|
|