The alarm package and modes of ventilation
including CPAP are similar to the IV100B infant ventilator
system.
This ventilator can adequately ventilate newborns to pediatric
patients
with precise control of oxygen concentration,
at respiratory rates from 1 to 150 BPM. To prevent inadvertent
PEEP when higher flow rates are used, a special pediatric
exhalation valve block and pediatric circuit are available.
Ventilation
Application |
|
Neonatal |
|
|
Pediatric
|
CONTROLS
|
|
|
Modes
of Ventilation |
|
OFF
|
|
|
CPAP |
|
|
VENT
(IMV,CMV) |
Flow |
|
0
to 32 LPM; Flush 40 LPM |
Oxygen
Percent |
|
21%
to 100% |
Inspiratory
Time |
|
0.10
to 2.90 sec. |
Expiratory
Time |
|
0.30
to 60.0 sec. |
Manual
Breath |
|
|
Inspiratory
Pressure |
|
5
to 70 cm H2O |
Expiratory
Pressure |
|
-2
to 20 cm H2O |
Test |
|
Checks
microprossor function, displays and alarm function |
|
|
Waveform
Control |
|
|
Allows
the user to vary the proximal airway pressure waveform |
Time
Preset |
|
While
in the CPAP mode, this feature allows the user to preview
and select the desired Inspiratory and Expiratory Times
before switching to the VENT mode. |
Safety
Pressure Relief Valve (Adjustable) |
|
15
to 85 cm H2O |
ALARM
CONTROLS |
|
|
Delay
Time |
|
3
to 60 sec. |
|
|
Sets
the time delay before the LO pressure alarm is audibly
and visually activated. |
Alarm
Mute |
|
30
± 5 sec. |
Alarm
Set Arrows |
|
High
Inspiratory Pressure Alarm Limit |
|
|
Low
Inspiratory Pressure Alarm Limit |
Select
(Reset) Button |
|
Allows
the user to select either the LO or HI pressure alarm
limit setting. This button is also used to cancel the
HI limit alarm when it is activated. |
Alarms |
|
High
(HI) Pressure Limit |
|
|
Peak
Overpressure |
|
|
Low
(LO) Pressure Limit |
|
|
Low
Airway Pressure |
|
|
Leaks |
|
|
Patient
Disconnect |
|
|
Fail-to-Cycle |
|
|
Apnea |
|
|
Prolonged
Inspiration |
|
|
Source
Gas Failure |
|
|
Power
Failure (Battery Backup) |
Displays |
|
Inspiratory
Time |
|
|
Expiratory
Time |
|
|
I:E
Ratio |
|
|
Rate |
|
|
Airway
Pressure (to 80 cm H2O) |
|
|
High
Pressure Alarm |
|
|
Low
Pressure Alarm |
|
|
Mean
Airway Pressure |
Indictators |
|
Inspiratory
Phase |
|
|
Inverse
I:E Ratio |
|
|
Mode
Selection |
|
|
Alarm |
|
|
High
and Low Pressure Alarm |
|
|
Limit
Set Points |
Pre-set |
|
Safety
Auto Lock-Out Circuit (Inspiratory Phase terminated
at 4 sec. Even if microprossor fails |
Pneumatic
Requirements |
|
Medical
Grade Air and Oxygen: 30 to 60 psi; differential within
20 psi. |
Electrical
Requirements |
|
117
VAC 50/60 Hz |
|
|
Overseas
voltage requirements available) |
|
|
Circuit
Breaker Protected |
|
|
D.C.
Power Requirements |
|
|
12
VDC 1.2 amps |
Physical
Specifications |
|
|
Dimensions
|
|
Height:
16.25 in (41.27 cm) |
|
|
Width:
12 in (30.48 cm) |
|
|
Depth:
9.5 in (24.13 cm) |
Weights |
|
Net:
27 lbs. (12.2 Kg) |
|
|
Shipping:
35 lbs. (15.9 Kg) |
Miscellaneous
Specifications |
|
3-Wire
SJT Power Cord |
|
|
Hospital-Grade
Plug, current leakage less than 20 microamps when tested
per UL 544 |
|
|
Auxiliary
Flowmeter Port |
|
|
*Specifications
subject to change without notice |
|
|
|